When carrying out the proving, the following measuring instruments are used:
1) pressure conversion channel:
- digital pressure gauge (pressure setter) of any type, measuring range up to 10 MPa, relative error not more than ± 0,05%;
2) temperature measuring channel:
- resistance store MCP-63, class. 0.05;
- thermostat “Thermotest-100”, temperature regulation range -30 ... + 100 ° С, variability of maintenance of set temperature ± 0,01 ° С, inhomogeneity of the temperature field in the working volume of the thermostat ± 0,01 ° С;
- the reference thermometer of resistance ETS-100, the third category (the range of measured temperatures is -50 ... + 419 ° С, the error is not more than ± 0,015 ° С;
- converter of signals "TERCON", bases measurement error of resistance ± [0,0002 + 1 × 10-5 × R], Ohm, voltage ± [0,0005 + 5 × 10-5 × Umeter] mV.
3) Flow measuring channel:
- pulse generator type G6-28, the basic error is not more than ± 1%;
- frequency meter Ч3-64 / 1, basic error is not more than ± 1.5. 10-7%;
4) general:
- an aneroid barometer M 67, measuring range from 81130 to 105320 Pa, error of not more than ± 106 Pa;
- hygrometer of psychrometric type VIT-1, range of measurement of relative humidity from 20 to 90%, measuring range of temperature from 15 to 40 °C, the division value of thermometer scales 0.1 °С.